Estudio piloto de una auditoría postal dosimétrica para radioterapia en condiciones de referencia

  1. Maria Carmen Pujades-Claumarchirant 1
  2. Cristian Candela-Juan 1
  3. Laura Oliver-Cañamás 1
  4. Ángela Soriano-Cruz 1
  5. Juan José Rovira-Escutia 1
  6. Facundo Ballester-Pallarés 2
  1. 1 Instituto Nacional de Gestión Sanitaria, España
  2. 2 Universidad de Valencia, España
Revista:
Revista de Física Médica

ISSN: 1576-6632

Any de publicació: 2022

Volum: 23

Número: 1

Pàgines: 27-34

Tipus: Article

DOI: 10.37004/SEFM/2022.23.1.002 DIALNET GOOGLE SCHOLAR lock_openAccés obert editor

Altres publicacions en: Revista de Física Médica

Resum

In 2019 a pilot study for postal dosimetry audits in radiotherapy centers was performed by the “Centro Nacional de Dosimetría”. The audit consisted in a verification of the calibration of megavoltage photon beams in reference conditions, for which optically stimulated luminescence (OSL) dosimeters placed in a water phantom were used. Forty-eight radiation beams were audited, with energies: 6, 10 and 15 MV with flattening filter and 6 and 10 MV flattening filter free (FFF). A difference between the declared absorbed dose and the measured dose below 5% was considered acceptable. The mean difference between the declared and measured doses was 0.8%, ranging from −0.9% to +1.9%. The uncertainty in the dose measurement was 1.6% (k = 1). No results above the acceptance limit were observed. From this study it is concluded that the proposed method to audit radiotherapy photon beams is viable and the results obtained for the different audited centers are below the acceptance limit. The uncertainty in the determination of the absorbed dose is similar to that of other organizations that offer this service.

Referències bibliogràfiques

  • Organismo Internacional de la Energía Atómica. IAEA TECDOC No. 1151 Aspectos físicos de la garantía de calidad en radioterapia: Protocolo de control de calidad. Viena, 2000. ISSN 1011-4289.
  • Belletti S, Dutreix A, Garavaglia G, Gfirtner H, Haywood J, Jessen KA, et al. Quality assurance in radiotherapy: the importance of medical physics staffing levels. Recommendations from an ESTRO/EFOMP joint task group. Radiother Oncol. 1996;41(1):89-94. https://doi.org/10.1016/s0167-8140(96)91799-5
  • Kouloulias VE. Quality assurance in radiotherapy. Eur J Cancer. 200;39(4):415-22. https://doi.org/10.1016/s0959-8049(02)00461-6
  • EURATOM. DIRECTIVA 2013/59/EURATOM del consejo de 5 de diciembre de 2013 por la que se establecen normas de seguridad básicas para la protección contra los peligros derivados de la exposición a radiaciones ionizantes, y se derogan las Directivas 89/618/Euratom, 90/641/Euratom, 96/29/Euratom, 97/43/Euratom y 2003/122/Euratom. Diario Oficial de la Unión Europea. 2013.
  • Organismo Internacional de la Energía Atómica. Quality Assurance Team for Radiation Oncology (QUATRO). Comprehensive Audits of Radiotherapy Practices: A Tool for Quality Improvement. Viena, 2007. ISBN 92–0–103707–4.
  • Kry SF, Peterson CB, Howell RM, Izewska J, Lye J, Clark CH, et al. Remote beam output audits: A global assessment of results out of tolerance. Phys and Im Rad Oncol. 2018;7:39-44. https://doi.org/10.1016/j.phro.2018.08.005
  • Izewska J, Bokulic T, Kazantsev P, Wesolowska P, van der Merwe D. 50 Years of the IAEA/WHO postal dose audit programme for radiotherapy: what can we learn from 13756 results? Acta Oncologica. 2020;59:459-502. https://doi.org/10.1080/0284186X.2020.1723162
  • Organismo Internacional de la Energía Atómica. Quality Assurance Team for Radiation Oncology (QUATRO) IAEA-TECDOC-1543. On-site Visits to Radiotherapy Centres: Medical Physics Procedures. Viena, 2007. ISBN 92–0–102607–2.
  • Alvarez P, Kry SF, Stingo F, Followill D. TLD and OSLD dosimetry systems for remote audits of radiotherapy external beam calibration. Radiat Meas. 2017;106:412-5. https://doi.org/10.1016/j.radmeas.2017.01.005
  • Kry SF, Alvarez P, Cygler JE, DeWerd LA, Howell RM, Meeks S. AAPM TG 191: Clinical use of luminescent dosimeters: TLDs and OSLDs. Med Phys. 2019;47:19-51. https://doi.org/10.1002/mp.13839
  • Wesolowska PE, Cole A, Santos T, Bokulic T, Kazantsev P, Izewska J. Characterization of three solid state dosimetry systems for use in high energy photon dosimetry audits in radiotherapy. Rad Meas. 2017;106:556-62. https://doi.org/10.1016/j.radmeas.2017.04.017
  • Izewska J, Andreo P. The IAEA/WHO TLD postal programme for radiotherapy hospitals. Radiot and Oncol. 2000;54:65-72. https://doi.org/10.1016/S0167-8140(99)00164-4
  • Derreumaux S, Chavaudra J, Bridier A, Rossetti V, Dutreix A. A European quality assurance network for radiotherapy: dose measurement procedure. Phys Med Biol. 1995;40(7):1191-208. https://doi.org/10.1088/0031-9155/40/7/004
  • Ferreira IH, Dutreix A, Bridierc A, Chavaudrac J, Svensson H. The ESTRO-QUALity assurance network (EQUAL). Radiot and Oncol. 2000;55:273-84. https://doi.org/10.1109/IEMBS.2000.901541
  • Okamoto H, Minemura T, Nakamura M, Mizuno H, Tohyama N, Nishio T, et al. Establishment of postal audit system in intensity-modulated radiotherapy by radiophotoluminescent glass dosimeters and a radiochromic film. Phys Med. 2018;48:119-126. https://10.1016/j.ejmp.2018.03.013
  • Angelis C, De Coste V, Fattibene P, Onori S, Petetti E. Use of alanine for dosimetry intercomparisons among Italian radiotherapy centers. App Rad Isot. 2005;62:261-5. https://doi.org/10.1016/j.apradiso.2004.08.019
  • Izewska J, Hultqvist M, Bera P. Analysis of uncertainties in the IAEA/WHO TLD postal dose audit system. Rad Meas. 2008;43:959-63. https://doi.org/10.1016/j.radmeas.2008.01.011
  • Izewska J, Georg D, Bera P, Thwaites D, Arib M, Saravi M, et al. A methodology for TLD postal dosimetry audit of high-energy radiotherapy photon beams in non-reference conditions. Radiot and Oncol. 2007;84:67-74. https://doi.org/10.1016/j.radonc.2007.06.006
  • Wesolowska P, Georg D, Lechner W, Kazantsev P, Bokulic T, Carlsson A, et al. Testing the methodology for a dosimetric end-to-end audit of IMRT/VMAT: results of IAEA multicentre and national studies. Acta Oncol. 2019;58(12):1731-9. https://doi.org/10.1080/0284186X.2019.1648859
  • Lechner W, Wesolowska P, Azangwe G, Arib M, Alves VGL, Suming L, et al. A multinational audit of small field output factors calculated by treatment planning systems used in radiotherapy. Phys and Im Rad Oncol. 2018; 5:58-63. https://doi.org/10.1016/j.phro.2018.02.005
  • Ibbott GS,Thwaites DI. Audits for advanced treatment dosimetry. J. Phys. Conf. Ser. 2015;573:012002. https://doi.org/10.1088/1742-6596/573/1/012002
  • Izewska J, Lechner W, Wesolowska P. Global availability of dosimetry audits in radiotherapy: The IAEA dosimetry audit networks database. Phys and Im Rad Oncol. 2018;5:1-4. https://doi.org/10.1016/j.phro.2017.12.002