Conductometria electronica. Un estudio comparativo

  1. PALLARES SABATER, ANTONIO
Supervised by:
  1. Vicente Faus Llácer Director

Defence university: Universitat de València

Year of defence: 1993

Committee:
  1. Jose V. Bagon Sebastian Chair
  2. Leopoldo Forner Navarro Secretary
  3. Carlos Canalda Sahli Committee member
  4. Rafael Romero Villafranca Committee member
  5. José María Vega del Barrio Committee member
Department:
  1. STOMATOLOGY

Type: Thesis

Teseo: 39419 DIALNET

Abstract

EL OBJETIVO DE LA PRESENTE TESIS DOCTORAL FUE COMPARAR LA PRECISION DE TRES LOCALIZADORES ELECTRONICOS DE APICE: 1) ODONTOMETER, DE USO CONVENCIONAL; 2) ENDOCATER, QUE SE USA CON UNAS SONDAS RECUBIERTAS DE MATERIAL AISLANTE EN TODA SU LONGITUD EXCEPTO EN LA PUNTA, CON EL FIN DE PODER MEDIR CONDUCTOS AUNQUE CONTENGAN ELECTROLITOS CONDUCTORES EN SU INTERIOR; 3) APIT, BASADO EN EL METODO DEL VALOR RELATIVO, QUE REQUIERE DE LA PRESENCIA DE ELECTROLITOS CONDUCTORES PARA PODER LLEVAR A CABO LA MEDICION. FUERON MEDIDOS UN TOTAL DE 168 CONDUCTOS PERTENECIENTES A 49 MOLARES. CUANDO EL CONDUCTO NO ESTABA SECO, CON EL ODONTOMETER, UN 79,3% DE LAS MEDICIONES QUEDARON EN LOS DOS INTERVALOS DE 0,5 MM MAS PROXIMOS A LA CONSTRICCION APICAL, MIENTRAS QUE CON EL ENDOCATER FUERON UN 88,7% Y CON EL APIT UN 90,4%. CON EL CONDUCTO SECO, CON EL ODONTOMETER QUEDARON EN LOS DOS INTERVALOS DE 0,5 MM MAS PROXIMOS A LA CONSTRICCION APICAL UN 84% DE LAS MEDICIONES Y UN 89,6% CON EL ENDOCATER.